My Techfocus

First time here? Sign Up


HYGROPHIL HCDT - Sample Conditioning System

  • NG # 003942-HW-WIN-O42

Manufactured by:

This seller generally responds in under 48 hours

The Sample Conditioning System Type 5987-72 (SCS - HCDT) is designed to condition gaseous media for determining the hydrogen and hydrocarbon dew point in conjunction with the gas composition analyzer Hygrophil F/HCDT.

Specification

MaterialsStainless Steel 1.4404 (and better)
Sample InletInlet Pressure: 500 kPa to 10 MPa (72.5 psi to 1450 psi)
Inlet Temperature: -20 °C to 60 °C (-4 °F to 140 °F)
Process Connection: 6 mm and 12 mm tube fittings (others on request)
Sample OutletOutlet Pressure: up to 380 kPa
Outlet Temperature: 5 °C to 40 °C (41 °F to 104 °F)
Electrical DataRated voltage: 230 VAC ± 10% 1Ph. 50 Hz
Rated current: approx. 4 A at 230 V without trace heating
Rated current: approx. 11 A at 230 V with trace heating
Power consumption: approx. 900 W (nominal)
Power consumption: approx. 2400 W (with trace heating)
DimensionsWidth: 915 mm
Height: 1070 mm
Depth: 300 mm
Weight: approx. 90 kg
Ambient conditionsStandard Temperature Range: 5 °C to 40 °C
Extended Temperature Range on request
Relative Humidity: 5 % to 100 % (non condensing / non corrosive)
Explosion ProtectionATEX II 2G Ex h IIC T3 Gb X
Ready for HydrogenHydrogen conent in Natural Gas ≤ 10 Mol %: on request
Hydrogen conent in Natural Gas ≤ 20 Mol %: on request
Operation in pure Hydrogen: on request

Description

The SCS - HCDT consists of various functional groups, which are custom designed and assembled for the relevant analysis task. The combination of individual components/assemblies does not result in any additional ignition hazards. The SCS - HCDT is suitable for operation in Zone 1.