HYGROPHIL HCDT - Sample Conditioning System
-
NG # 003942-HW-WIN-O42
Manufactured by:
The Sample Conditioning System Type 5987-72 (SCS - HCDT) is designed to condition gaseous media for determining the hydrogen and hydrocarbon dew point in conjunction with the gas composition analyzer Hygrophil F/HCDT.
Specification
Materials Stainless Steel 1.4404 (and better) Sample Inlet Inlet Pressure: 500 kPa to 10 MPa (72.5 psi to 1450 psi)
Inlet Temperature: -20 °C to 60 °C (-4 °F to 140 °F)
Process Connection: 6 mm and 12 mm tube fittings (others on request)Sample Outlet Outlet Pressure: up to 380 kPa
Outlet Temperature: 5 °C to 40 °C (41 °F to 104 °F)Electrical Data Rated voltage: 230 VAC ± 10% 1Ph. 50 Hz
Rated current: approx. 4 A at 230 V without trace heating
Rated current: approx. 11 A at 230 V with trace heating
Power consumption: approx. 900 W (nominal)
Power consumption: approx. 2400 W (with trace heating)Dimensions Width: 915 mm
Height: 1070 mm
Depth: 300 mm
Weight: approx. 90 kgAmbient conditions Standard Temperature Range: 5 °C to 40 °C
Extended Temperature Range on request
Relative Humidity: 5 % to 100 % (non condensing / non corrosive)Explosion Protection ATEX II 2G Ex h IIC T3 Gb X Ready for Hydrogen Hydrogen conent in Natural Gas ≤ 10 Mol %: on request
Hydrogen conent in Natural Gas ≤ 20 Mol %: on request
Operation in pure Hydrogen: on request
Description
The SCS - HCDT consists of various functional groups, which are custom designed and assembled for the relevant analysis task. The combination of individual components/assemblies does not result in any additional ignition hazards. The SCS - HCDT is suitable for operation in Zone 1.